    1. SPECTRAL SLIT WIDTH (if applicable)
    When using an instrument on which the slit-width is variable at the selected wavelength, the
    slit-width must be small compared with the half-width of the absorption band but it must be
    as large as possible to obtain a high value of I0. Therefore, a slit-width is chosen such that
    further reduction does not result in a change in absorbance reading.
    Method and Limits:
    1. Switch the system on and start the Scan module.
    2. Select SETUP and set the following parameters:
    X Mode = Nanometres
    Start wavelength = 660.0 nm
    Stop wavelength = 650.0 nm
    Y Mode = % T
    Scan rate = 10 nm/min
    Gain = (100) see 4 below
    PA/PH/OMCL (07) 11 DEF CORR - OMCL Guideline on Qualification of UV-Visible spectrophotometers (Annex 3)
    ···